The 236, 237, and 238 Source-Measure Units (SMU) are fully programmable instruments, capable of sourcing and measuring voltage or current simultaneously. These systems are really four instruments in one: voltage source, current source, voltage measure and current measure.

The 236 will source voltage from 100µV to 110V, and current from 100fA to 100mA. It can also measure voltage from 10µV to 110V and current from 10fA to 100mA. The 237 offers the same capabilities with a decade enhancement in voltage source and measure (1100V). In this higher voltage range, current source and measure is 10mA maximum. The 238 offers a decade enhancement in current source and measure (1A). In this higher current range, voltage source and measure is 15V maximum.

  • Load Capacitance: Stable into 20,000pF typical.
  • Remote Sense: Corrects for up to 2V drop in each output lead. Maximum 1kΩ per sense lead for rated accuracy. Residual output resistance (as a voltage source) is 0.5Ω.
  • Guard: Output Resistance: ≤12kΩ.
  • Maximum Output Current: ±2mA.
  • Offset Relative to Output HI: ±2mV max.
  • Isolation (Output LO to chassis): Typically >1010Ω in parallel with 500pF.
  • Maximum Common Mode Voltage: 200V.
  • Connector: Outputs: 3-lug triax.
  • Trigger Input/Output: BNC.
  • Interlock: 3-pin miniature DIN.
  • Temperature Coefficient (0°–18°C and 28°–50°C): ±(0.1 × applicable accuracy specification)/°C.
  • Environment: Operating: 0°–50°C, 70% relative humidity up to 35°C. Linearly derate 3% R.H./°C, 35°–50°C, Storage: –25° to 65°C.
  • EMC: Conforms to European Union Directive 89/336/EEC.
Description

Keithley 237 High Voltage Source Measure Unit

The 236, 237, and 238 Source-Measure Units (SMU) are fully programmable instruments, capable of sourcing and measuring voltage or current simultaneously. These systems are really four instruments in one: voltage source, current source, voltage measure and current measure.

The 236 will source voltage from 100µV to 110V, and current from 100fA to 100mA. It can also measure voltage from 10µV to 110V and current from 10fA to 100mA. The 237 offers the same capabilities with a decade enhancement in voltage source and measure (1100V). In this higher voltage range, current source and measure is 10mA maximum. The 238 offers a decade enhancement in current source and measure (1A). In this higher current range, voltage source and measure is 15V maximum.