Haefely PET 4000 (249997) ESD Calibration Set

Manufacturer part number: PET 4000
ESD Calibration Set for IEC/EN 61000-4-2
PURCHASE
AvailableQTYPrice
New:   0
+ -
$0.00
Used:   0
+ -
$0.00Configurations
X

Available Configurations

ESD generators are calibrated according to the standard IEC/EN 61000-4-2 Ed 2.0. The Haefely PET 4000 ESD Current Target Calibration Set allows the user to take measurements conforming to the new 4GHz requirement in accordance with IEC 61000-4-2 Ed. 2.0.

Measurement of individual impulse parameters is performed using a special coaxial current shunt. This has a nominal impedance of 2 Ohms. Because ESD events can be generated as high as 120A, attenuators are required to reduce the impulse-voltage level so it can be displayed on commercial oscilloscopes. An oscilloscope with minimum 2GHz bandwidth is required to properly display the ESD impulse.

Variation of Insertion Loss :+/- 0.5 dB up to 1 GHz, +/- 1.2 dB up to 4 GHz,

Input/Shunt Resistance: 2 Ohm

Pulse Amplitude: 30kV / 120A Maximum

Bandwidth: 4 GHz

  • ESD target 2 Ohm / 4GHz
  • One 20dB / 18GHz attenuator (additional 20dB attenuator optionally, ordering number 0780841)
  • Coaxial cable - 1 meter
  • User manual
Description

ESD generators are calibrated according to the standard IEC/EN 61000-4-2 Ed 2.0. The Haefely PET 4000 ESD Current Target Calibration Set allows the user to take measurements conforming to the new 4GHz requirement in accordance with IEC 61000-4-2 Ed. 2.0.

Measurement of individual impulse parameters is performed using a special coaxial current shunt. This has a nominal impedance of 2 Ohms. Because ESD events can be generated as high as 120A, attenuators are required to reduce the impulse-voltage level so it can be displayed on commercial oscilloscopes. An oscilloscope with minimum 2GHz bandwidth is required to properly display the ESD impulse.