Test Level CWS 500N1.3
|
Output Level
|
1 - 30Ve.m.f. after 6dB-attenuator
|
Output power
|
80W (nominal)
|
Output impedance
|
50 Ohm
|
max. VSWR
|
1:1.2 at all phase angles and at max. power (without destruction)
|
Harmonic distortion (at max. power)
|
< 15dBc
|
Test frequencies
|
CWS 500N1.3
|
CWS 500N1.4
|
Sinusoidal (CW)
|
10 kHz – 400 MHz
|
100 kHz – 300 MHz step: 1kHz
|
|
In the menu QUICK START the step size can be selected by the user
|
Modulation
|
Modulation method
|
Amplitude modulation 80% <±5% and 1kHz <±10% acc. to IEC 61000-4-6 80% <±5% and 2Hz, 1kHz acc. to IEC 60601-1-2
|
Pulse modulation
|
1Hz, 50% duty cycle acc. to EN 50130-4
|
General
|
80% at Frequencies 2Hz, 400Hz,1kHz
|
Other modulation
|
via ICD software Settings in Vector mode
|
Signal Generator
|
Output level
|
-63.5 dBm to 0 dBm
|
Step level
|
0.5 dB
|
Frequency range
|
9 kHz to 1 GHz
|
Output impedance
|
50 ohm
|
Calibration measurements
|
Cal data
|
5 internal stores for calibration data.
|
ICD
|
Software for the upload and download of Cal-files.
|
LCD
|
On-line display of the test level and the preselected frequencies.
|
Monitor input PM1000
|
RF monitor input for calibration procedures and current probes measurement. Input voltage for current monitoring V max = 1.0 V rms (13dBm)
|
Cal data for current probe (CP)
|
2 internal stores ; ( download with ICD software ) A download will reduce data to 10/ decade and frequency range 100kHz ... 300MHz
|
Timing
|
Dwell time
|
general td = 0,3s - 9999.9s pulse modulation td = 3s - 9999.9s
|
Rest time
|
tr = 0 / 0,3s - 9999.9s
|
Output
|
Direct
|
N
|
EUT control
|
BNC input
|
FAIL 1 Fail 1; test will be stopped immediately
|
BNC input
|
FAIL 2 Fail 2; failure is detected, test continues (max. 10 failures)
|
Test routines (integrated)
|
Quick Start
|
Immediate start, all parameters adjustable during testing.
|
User test routines
|
1. Voltage sweep 2. Frequency sweep 3. Dwell time sweep
|
Standard test routines.
|
1. Level 1-3 acc. to IEC 61000-4-6 2. Automatic Level X - Level Y
|
Cal procedure
|
Calibration of the complete test set-up. Internal storage of the Cal-files.
|
Service
|
Service, Setup
|
Interfaces
|
Serial Interface
|
USB
|
Parallel IEEE 488 GPIB interface
|
Addresses 1 - 30 selectable All interfaces are included as standard features.
|
Power meter PM 1000
|
Frequency range
|
9kHz – 1000 MHz
|
Input range
|
Monitor -45 dBm ... +13 dBm FWD power -50 dBm ... +13 dBm REV power -50 dBm ... +13 dBm
|
Max. Input MONITOR
|
13dBm approx. 1.0 Vrms
|
Accuracy
|
< ± 0.5 dB ( 0.09MHz...400MHz ) < ± 1.0 dB ( 0.01MHz...1000MHz )
|
Dual Directional Coupler
|
Frequency range
|
10kHz – 1000 MHz
|
Power
|
150 W max.
|
Insertion loss
|
0.6 dB max.
|
Mainline VSWR
|
1.1:1 max.
|
RF Amplifier ( build in )
|
CWS 500N1.3
|
CWS 500N1.4
|
Class of Operation
|
Class A
|
Class A
|
Frequency range
|
10 kHz .. 400 MHz
|
100 kHz .. 300 MHz
|
Output power nominal
|
80 W
|
80 W
|
Gain
|
48 dB nominal
|
48 dB nominal
|
Input power for rated output
|
0 dBm
|
0 dBm
|
Input power max.
|
10 dBm
|
10 dBm
|
Input / Output Impedance
|
50 Ω
|
50 Ω
|
Input VSWR
|
1.5: 1 max.
|
1.5: 1 max.
|
General Data
|
Dimensions
|
19"/3HE
|
19"/3HE
|
Weight
|
17.25 kg
|
14.95 kg
|
Power supply
|
110-230V/ max 50/60Hz
|
Input power
|
max. 380W
|
Power factor
|
cosΦ=0,98 at max output power
|
acc. to IEC 555
|
Fuse
|
230V: 2x3,15AT 115V: 2x6.3AT
|
Cooling
|
Active cooling, air ventilation
|
Environment conditions
|
10°C - 35°C
|
10°C - 40°C
|
Humidity
|
Max. 85 %, non-condensing
|