Magnetic Field Adjustment: Manually via potentiometer
Range Low: 80 to 400 mA into MFC 30 (Cf 9,8) allows 0,8 to 4 A/m; 80 to 400 mA into MFC 300 (Cf 34) allows 2,7 to 13,6 A/m
Range High: 200 to 4,1 A into MFC 30 (Cf 9,8) allows 2 to 40 A/m; 200 to 4,1 A into MFC 300 (Cf 34) allows 6,8 to 130 A/m
Supply Voltage: Wide-range power-supply 90 to 240 VAC
Power Consumption: < 150 W
Operating Temperature: 5° - 40° C (41° - 104° F)
Overload Protection: By temperature sensor on power stage
Total Harmonic Distorsion (THD): < 8% (nominal < 3,5% at full range)
Frequency: Selectable 50 and 60 Hz +/- 3%
Operating Position: Laying or standing on a work bench, or for more permanent applications, it can be wall mounted
Suitable Loads: MFG 40-100 is designed to drive INDUCTIVE LOADS ONLY, such as multi-turn magnetic field loops. Connecting capacitive loads will destroy the instrument
Dimensions: 195 x 180 x 380 mm (7.67 x 7.08 x 14.96 in)
Weight: 4 kg (8.81 lbs)