The ACT100 & DCT350 are engineered to provide unparalleled accuracy and efficiency in EVSE testing. Built for laboratory and manufacturing environments, these rack-mounted systems provide real-time waveform generation, harmonic analysis, and automated calibration to ensure accuracy across AC and DC electric vehicle charging systems.
Key Features :
Compliance testing of electrical and electronic devices
Functional and endurance testing of power components
Insulation and dielectric testing
Production line verification
High-voltage and high-current safety tests
R&D lab validation of energy and power systems
ACT-100
Digital Waveform Generator Capabilities
Three voltage and three current channels
Synthesizer
Fundamental frequency adjustable from 45 to 65 Hz
0.01 Hz resolution
Sinusoidal (THD < 0.4%) (< 0.1% typical)
ANSI harmonics and custom harmonics
Current
0.1A – 100A rms
0.01A adjustment increment
0° – 360° phase
0.001° phase adjustment increment
Accuracy: 0.2%
Compliance voltage: 1V
Open circuit protection
Voltage
30V – 350V Line to Neutral
0V – 700V Line to Line
0.01V adjustment increment
Output power: 100VA
Short circuit protection
Integrated Reference Standard
Measurement accuracy (60 Vac – 350 Vac, 0.25 Aac – 80 Aac @ 60Hz, 1.0 PF):
Phase: ±0.01°
Real Energy: ±0.04%
Apparent Energy: ±0.04%
Reactive Energy: ±0.04%
Measures to the 50th harmonic
Time domain and frequency domain measurements
PHYSICAL
Size: 25.54”L x 25.62”W x 75.5”H
15” shelf
21” screen
Main power break
Cabling
ACT-100 – DB9 pulse cable
ACT-100 – CCS1 AC calibration cable
ACT-100 – NACS AC calibration cable
ACT-100 – AC return cable
ACT-100 – External reference calibration cable
INPUT
120V, 8A, 60Hz
Ethernet cable for network connectivity
DB9 header for calibration and synchronization with device under test
CALIBRATION INTERFACE
Connectors for calibration and return cabling
RJ45 jack for communication with device under test
DCT-350
Two voltage and two current channels
1V – 1000V Line to Line, center tapped to GND
1V adjustment increment
Accuracy: 1%
1A – 630A rms
1A adjustment increment
Compliance voltage: 2V
Measurement accuracy (100 Vdc – 1000 Vdc, 20 Adc – 600 Adc):
Voltage: ±0.05%
Current: ±0.05%
Energy: ±0.1%
DCT-350 – DB9 pulse cable
DCT-350 – CCS1 DC calibration cable
DCT-350 – NACS DC calibration cable
DCT-350 – DC return cable
240V, 30A, 60Hz
Spec sheet ACT-100
Spec sheet DCT-350